[. . . ] [. . . ] All PTB100 barometers are delivered with a factory calibration certificate, which is NIST traceable.
Excellent long-term stability
The PTB100 barometer is designed both for accurate barometric measurements at room temperature and for general environmental pressure monitoring over a wide temperature range. The excellent long-term stability of the barometers minimizes or even removes the need for field adjustment in many applications.
Ideal for data logger applications
The compact PTB100 barometer is ideal for data logger applications because of the low power consumption, selectable external on/off control, practical output voltage ranges and three or four wire connection capability.
112
Vaisala Instruments Catalog 2007
Ref. A
PTB100
BAROMETRIC PRESSURE
Technical Data
Operating range
Pressure range PTB100A PTB100B/PTB 101B PTB101C Temperature range Humidity range
(1hPa=1mbar)
800. . . 1060 hPa 600. . . 1060 hPa 900. . . 1100 hPa -40. . . +60 °C (-40. . . +140 °F) non-condensing
Dimensions
Dimensions in mm (inches)
Accuracy
PTB100A/PTB101C PTB100B/PTB101B Linearity* ±0. 25 hPa ±0. 45 hPa Hysteresis* ±0. 03 hPa ±0. 05 hPa Repeatability* ±0. 03 hPa ±0. 05 hPa Calibration uncertainty** ±0. 15 hPa ±0. 15 hPa Accuracy at +20 °C (+68 °F)*** ±0. 3 hPa ±0. 50 hPa * Defined as ±2 standard deviation limits of end point non-linearity, hysteresis error or repeatability error ** Defined as ±2 standard deviation limits of inaccuracy of the working standard at 1000 hPa including traceability to NIST ***Defined as the root sum of the squares (RSS) of end-point non-linearity, hysteresis error, repeatability error and calibration uncertainty at room temperature Total accuracy +20 °C (+68 °F) 0. . . +40 °C (+32. . . 104 °F) -20. . . +45 °C (-4. . . 113 °F) -40. . . +60 °C (-40. . . 140 °F) PTB100A/PTB101C PTB100B/PTB101B ±0. 3 hPa ±0. 5 hPa ±1. 0 hPa ±1. 5 hPa ±1. 5 hPa ±2. 0 hPa ±2. 5 hPa ±3. 0 hPa ±0. 1 hPa/year < ±0. 2 hPa
21. 8 (0. 86)
60 (2. 36)
4 (0. 16) 76. 2 (3. 00)
3. 5 (0. 14)
Long-term stability Effect of thermal or mechanical shocks
General
Supply voltage 10. . . 30 VDC Supply voltage control with TTL level trigger when enabled with an internal jumper, barometer can be triggered on/off using external TTL level trigger Supply voltage sensitivity less than 0. 1 hPa Current consumption operation mode less than 4 mA shutdown mode less than 1 µA Output voltage PTB100A/PTB 100B 0. . . 5 VDC PTB101B/PTB101C 0. . . 2. 5 VDC Resolution 0. 1 hPa Load resistance 10 kohm minimum Load capacitance 47 nF maximum Settling time at power-up 1s Response time (100% response) 300 ms Warm-up shift less than 0. 1 hPa Acceleration sensitivity negligible Pressure connector M5 (10-32) internal thread Pressure fitting barbed fitting for 1/8" I. D. Electrical connector a removable connector for five wires (AWG 28. . . 16) Housing material aluminum Weight 85 g Electromagnetic compatibility Complies with EMC standard EN61326-1, Generic Environment
13 (0. 51)
96. 9 (3. 81)
11. 5 (0. 45)
7. 2 (0. 28)
BAROCAP® is a registered trademark of Vaisala. [. . . ] [. . . ]