User manual VAISALA HMT310
DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!
If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual VAISALA HMT310. We hope that this VAISALA HMT310 user guide will be useful to you.
You may also download the following manuals related to this product:
VAISALA HMT310 DATASHEET (429 ko)
Manual abstract: user guide VAISALA HMT310
Detailed instructions for use are in the User's Guide.
[. . . ] USER'S GUIDE
Vaisala HUMICAP® Humidity and Temperature Transmitter HMT310
M210619EN-C
PUBLISHED BY Vaisala Oyj P. O. Box 26 FIN-00421 Helsinki Finland Phone (int. ): +358 9 8949 1 Fax: +358 9 8949 2227
Visit our Internet pages at http://www. vaisala. com/ © Vaisala 2007 No part of this manual may be reproduced in any form or by any means, electronic or mechanical (including photocopying), nor may its contents be communicated to a third party without prior written permission of the copyright holder. The contents are subject to change without prior notice.
Please observe that this manual does not create any legally binding obligations for Vaisala towards the customer or end user. All legally binding commitments and agreements are included exclusively in the applicable supply contract or Conditions of Sale.
________________________________________________________________________________
Table of Contents
CHAPTER 1
GENERAL INFORMATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . [. . . ] The unit is calibrated at the factory and the device is ready for use. The transmitter is delivered with a detachable 5 m cable with eight wires for serial port, analog outputs and 24VDC power supply or with a screw terminal connector. See the wiring instructions in Figure 13 below and in Figure 14 on page 29.
Cable Wiring
0507-044
Figure 13
8-Pole Connector
28 _______________________________________________________________ M210619EN-C
Chapter 3 _______________________________________________________________ Installation
The following numbers and codes refer to Figure 13: Color 1 White (WHT) 2 Brown (BRN) 3 Green (GRN) 4 Yellow (YEL) 5 Grey (GREY) 6 Pink (PINK) 7 Blue (BLU) 8 Red (RED) = = = = = = = = Function RS-232 TX RS-232 GND CH2+ CH1+ Supply-/CH1-/CH2 Supply+ RS-232 RX Not connected
Screw Terminal Connector
0507-045
Figure 14
Screw Terminal Connector
The following numbers refer to Figure 14 above: Color 1 White (WHT) 2 Brown (BRN) 3 Green (GRN) 4 Yellow (YEL) 5 Grey (GREY) 6 Pink (PINK) 7 Blue (BLU) 8 Red (RED) = = = = = = = = Function RS-232 TX RS-232 GND CH2+ CH1+ Supply-/CH1-/CH2 Supply+ RS-232 RX Not connected
VAISALA____________________________________________________________________ 29
USER'S GUIDE____________________________________________________________________
This page intentionally left blank.
30 _______________________________________________________________ M210619EN-C
Chapter 4 ________________________________________________________________ Operation
CHAPTER 4
OPERATION
This chapter contains information that is needed to operate Humidity Transmitter HMT310.
Power ON/OFF
Switch on the 24 VDC power supply and the transmitter wakes up.
Entering Serial Communication Parameters
The transmitter communicates via an RS-232 serial interface. The transmitter can be polled or set on RUN-mode with specific commands. The data format will be (factory settings): - 1 start bit - 7 data bits - 1 stop bit - Even parity - 4800 bits per second, programmable to 19200 - Full duplex - Serial asynchronous - Configured as Data Terminal Equipment (DTE) Enter the communication parameters when using this terminal session for the first time; save them for future use. See instructions in Table 4 on page 32.
VAISALA____________________________________________________________________ 31
USER'S GUIDE____________________________________________________________________
Table 4
WINDOWS 2000
MENU WHAT TO DO
Entering Serial Parameters in Windows® 2000 and Windows NT®
WINDOWS NT
MENU WHAT TO DO
Start move the cursor to: Programs move the cursor to: Accessories move the cursor to: Communications move the cursor to: HyperTerminal Hypertrm click move the cursor to: double click type the name of the connection in the appropriate field and select an icon if available; click OK. move the cursor to the field CONNECT USING and select 'direct to COM x' (x = serial port available); click OK select parameters according to the following figure; click OK
Start move the cursor to: Programs move the cursor to: Accessories move the cursor to:
HyperTerminal move the cursor to: Hyperterminal click type the name of the connection in the appropriate field and select an icon if available; click OK. move the cursor to the field CONNECT USING and select 'COM x' (x = serial port available); click OK select parameters according to the following figure, click OK
Connection Description
Connection Description
Phone Number
Connect to
COM x properties
COM x properties
32 _______________________________________________________________ M210619EN-C
Chapter 4 ________________________________________________________________ Operation
0507-046
Figure 15
HyperTerminal Settings in Windows 2000 Environment
®
List of Commands
The text in BOLD letters in the [brackets] indicates the default setting. This section presents the command list, the commands are described in detail later in this chapter. 99] CLOSE Start continuous outputting Stop continuous outputting Set continuous output interval for RUN-mode Output reading once Set serial interface mode Serial line settings (default: 4800 E 7 1) baud: 300. . . 19200 Set transmitter address for Use in POLL-mode Temporarily open transmitter from POLL-mode to receive commands Set transmitter in POLL-mode
VAISALA____________________________________________________________________ 33
USER'S GUIDE____________________________________________________________________
Output formatting FORM TIME DATE FTIME [ON/OFF] FDATE [ON/OFF] UNIT FST [ON/OFF]
Serial output format Set time Set date Add time to SEND and R outputs Add date to SEND and R outputs Select metric or nonmetric output units Output state of optional probe heating and chemical purge (with SEND and R commands) Check transmitter settings Check transmitter settings in POLL-mode Serial bus echo All devices in POLL-mode send their addresses List commands Set ambient pressure for calculations (non-volatile) Set ambient pressure for calculations Set result filtering Reset transmitter Display error messages
Other commands ?ECHO [ON/OFF] FIND HELP PRES XPRES FILT RESET ERRS
Setting, scaling and testing analog outputs AMODE Set analog outputs (0/4 . . . 20 mA) ASEL Select parameter for analog outputs ASCL Scale analog outputs ITEST Test analog outputs AQTEST Test analog outputs for desired readings AERR Set error outputs
34 _______________________________________________________________ M210619EN-C
Chapter 4 ________________________________________________________________ Operation
Calibration and adjustment (these are presented in Chapter 7, section Calibration and Adjustment Commands on page 64) CRH Relative humidity calibration CT Temperature calibration LI Revert factory calibration FCRH Relative humidity calibration after sensor change CTEXT Set calibration information text CDATE Set calibration date ACAL Analog output calibration Chemical purge PUR [ON/OFF] PURGE PURR [ON/OFF] Enable or disable automatic chemical purge Activate chemical purge Enable or disable automatic chemical purge in power up Enable or disable sensor heating
Sensor heating XHEAT [ON/OFF]
Measurement Output R Start Continuous Outputting
Syntax: R<cr> Starts output of measurements to the peripheral devices (RUN-mode); the only command that can be used is S (stop). Example:
>r<cr> RH= 28. 0 RH= 28. 0 RH= 28. 0 RH= 28. 0 RH= 28. 0 RH= 28. 0 %RH %RH %RH %RH %RH %RH T= T= T= T= T= T= 23. 3 23. 3 23. 3 23. 3 23. 3 23. 3 'C 'C 'C 'C 'C 'C
VAISALA____________________________________________________________________ 35
USER'S GUIDE____________________________________________________________________
S Stop Continuous Outputting
Syntax: S<cr> Stops the continuous output. Also the Esc key (on your computer keyboard) can be used to stop outputting.
INTV Set Continuous Output Interval for RUN-Mode
Syntax: INTV xxx yyy<cr> where xxx yyy Example:
>intv 1<cr> Output interval: 1 S >intv 1 min<cr> Output interval: 1 MIN >intv 1 h<cr> Output interval: 1 H
= =
Output interval (0 . . . 255) Unit (s, min or h)
SEND Output Reading Once
Syntax in STOP-mode: SEND<cr> Syntax in POLL-mode: SEND aa<cr> where aa = Address of the transmitter when more than one transmitter is connected to a serial bus (0 . . . 99)
36 _______________________________________________________________ M210619EN-C
Chapter 4 ________________________________________________________________ Operation
SMODE Set Serial Interface Mode
Syntax: SMODE x<cr> where x = STOP/RUN/POLL STOP-mode: Transmitter in standby for serial commands RUN-mode: Transmitter outputs data continuously POLL-mode: Transmitter only responds to addressed command
Example:
>smode run<cr> Output mode : RUN >smode stop<cr> Output mode : STOP
SERI Serial Line Settings
Syntax: SERI b p d s<cr> where b p d s = = = = bauds (300, 600, 1200, 2400, 4800, 9600, 19200) parity (n = none, e = even, o = odd) data bits (7 or 8) stop bits (1 or 2)
The settings can be changed one parameter at a time or all parameters at once: Example:
>seri o<cr> 4800 o 7 1 >seri 600 n 8 1<cr> 600 n 8 1
changing parity only changing all parameters
VAISALA____________________________________________________________________ 37
USER'S GUIDE____________________________________________________________________
ADDR Set Transmitter Address for Use in POLL-Mode
Syntax: ADDR aa<cr> where aa Example:
>addr<cr> Address >addr 1<cr> Address :0 :1
=
Address (0 . . . 99)
OPEN Temporarily Open Transmitter from POLL-Mode to Receive Serial Commands
Syntax: OPEN nn<cr> where nn = Address of the transmitter (0 . . . 99)
The OPEN command sets the bus temporarily in STOP-mode so that the SMODE command can be issued. Example:
>open 4<cr> Device: 4 line opened for operator commands >
38 _______________________________________________________________ M210619EN-C
Chapter 4 ________________________________________________________________ Operation
CLOSE Set Transmitter in POLLMode
Syntax: CLOSE<cr> In STOP-mode: command OPEN has no effect, CLOSE sets the transmitter temporarily in POLL-mode. In POLL-mode: command OPEN sets the transmitter temporarily in STOP-mode, command CLOSE returns the instrument to POLLmode. Example: Relative humidity calibration is performed at transmitter 2, which is in POLL-mode.
>open 2<cr>
Opens the line to transmitter 2.
>crh<cr>
Calibration started.
. . . >close<cr>
Line closed.
Output Formatting FORM Serial Output Format
Syntax: FORM x<cr> where x = Formatter string
Command format can be used to change the format of the output commands SEND and R.
VAISALA____________________________________________________________________ 39
USER'S GUIDE____________________________________________________________________
Format string consists of quantities and modifiers: use the quantity abbreviations and modifiers presented in Table 5 below and in Table 6 below when selecting the output quantities. [. . . ] +180 °C Factory calibration uncertainty (+20 °C)
for typical applications for typical applications for applications with chemical purge and/or warmed probe for applications with chemical purge and/or warmed probe ± 1 % RH (0 . . . 100 %RH) ± (1. 0 + 0. 008 × reading) %RH ± (1. 5 + 0. 015 × reading) %RH for applications with demanding chemical environment ± (1. 0 + 0. 01 × reading) %RH ± (1. 5 + 0. 02 × reading) %RH ±0. 6 % RH (0 . . . Small variations possible, see also calibration certificate. ) 8 s with grid filter 20 s with grid + steel netting filter 40 s with sintered filter
Response time (90 %) for HUMICAP®180 and HUMICAP®180C at 20 °C in still air
72 _______________________________________________________________ M210619EN-C
Chapter 8 ____________________________________________________________ Technical Data
Response time (90 %) for HUMICAP®180R and HUMICAP®180RC at 20 °C in 0. 1 m/s air flow
17 s with grid filter 50 s with grid + steel netting filter 60 s with sintered filter
Table 12
Temperature Specifications (+ Operating Pressure Ranges)
Description / Value -40 . . . 40 bar) ±0. 2 °C See graph below Pt 100 RTD 1/3 Class B IEC 751
Property HMT311 HMT313 80 °C HMT313 120 °C HMT314 HMT315 (vapor tight) HMT317 (vapor tight) HMT318 Accuracy at +20 °C (+68 °F) Accuracy over temperature range Temperature sensor
0. 7 0. 6
°C
0. 5 0. 4 0. 3 0. 2 0. 1 0 -0. 1 -0. 2 -0. 3 -0. 4 -0. 5 -0. 6 -0. 7 -80 -60 -40 -20 0 20 40 60 80 100 120 140 160 180
0507-021
°C
Figure 18
Accuracy over Temperature Range
VAISALA____________________________________________________________________ 73
USER'S GUIDE____________________________________________________________________
Table 13
Electronics Specifications
Description / Value 8-pole connector with 5 m cable Female 8-pin connector screw joint for cable diameter 4 . . . [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE VAISALA HMT310
Click on "Download the user Manual" at the end of this Contract if you accept its terms, the downloading of the manual VAISALA HMT310 will begin.