User manual VAISALA GMW115 QUICK REFERENCE GUIDE
DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!
If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual VAISALA GMW115. We hope that this VAISALA GMW115 user guide will be useful to you.
You may also download the following manuals related to this product:
VAISALA GMW115 DATASHEET (371 ko)
Manual abstract: user guide VAISALA GMW115QUICK REFERENCE GUIDE
Detailed instructions for use are in the User's Guide.
[. . . ] 10 V Signal (-) RS-485 Signal B RS-485 Signal A Power supply (-) Power supply (+) 24 VDC/VAC
0601-002
Figure 2 Connection of Single AC supply to Several Transmitters
MOUNTING AND DIMENSIONS
Dimensions
mA V 0 B A 0 24V mA V 0 B A 0 24V
QUICK REFERENCE GUIDE
See Figure 3 for wire terminals. Powering
25
Vaisala CARBOCAP® Carbon Dioxide Module GMM112 and Transmitter GMW115
The products require a nominal 24 VDC/VAC power supply maintaining a voltage of 18 . . . [. . . ] Make sure the back plate is mounted in with the same orientation as in Figure 4. GMW115 Cover Opening/Closing
SERIAL COMMANDS
Serial commands are the same for RS232 and RS485 interface. Polling command for CO2 measurement (ppm): SEND <cr> Setting the interval for the RUN (continuous output) mode: INTV X Y<cr> X = 0 (default) . . . 255 Y = S/MIN/H Starting the continuous mode printing: R<cr> Stopping the continuous mode printing: S<cr> Saving the parameters into the memory: SAVE<cr> Enabling/disabling the RS485 interface: RS485 X<cr> X = ON/OFF Changing the operation mode: SMODE X<cr> X = STOP (default) / RUN / POLL Giving the device address: ADDR X<cr> X = 0 (default) . . . 99 Opening the polling line: OPEN addr<cr> addr = 0 (default) . . . 99 Closing the polling line: CLOSE<cr>
0601-007
TECHNICAL DATA
Property Performance Measurement ranges Measurement accuracy (incl. repeatability, nonlinearity and calibration uncertainty) Long-term stability Response time Temperature dependence of reading Pressure dependence of reading Warm-up time Product lifetime Operating environment Operating temperature range Operating humidity range Operating pressure range Inputs and outputs Operating voltage Power consumption Connections Outputs analog serial Recommended external load current output voltage output Electromagnetic compatibility Materials Weight Housing material Description / Value 0 . . . 5000 ppm CO2 ±(2 % of range + 2 % of reading)
0601-006
± 5 % of range / 5 years 1 min -0. 35 % of reading / °C (typical) + 0. 15 % of reading / hPa (typical) 1 min 10 min full specification > 10 years -5 . . . 1200 hPa 24 V (±20 %) AC/DC <2W Screw terminals, wire size 2 0. 5 . . . 20 mA RS-485, 2-wire, non-isolated
Figure 5 Opening the GMW115 Cover
SERVICE, CALIBRATION AND ADJUSTMENT
These products are designed to operate their lifetime without maintenance The transmitters are calibrated before shipping from the factory. For technical questions, contact the Vaisala technical support; by e-mail helpdesk@vaisala. com, or by fax +358 9 8949 2790. Repair and calibration services are provided by Vaisala Service Centers:
Figure 6 Closing the GMW115 Cover
SERIAL COMMUNICATION INTERFACES
The modules support RS232 or RS485 communication. [. . . ] Visit our Internet pages at
GUARANTEE Vaisala issues a guarantee for the material and workmanship of this product under normal operating conditions for two (2) years from the date of delivery. Exceptional operating conditions, damage due to careless handling and misapplication will void the guarantee. [. . . ]
DISCLAIMER TO DOWNLOAD THE USER GUIDE VAISALA GMW115
Click on "Download the user Manual" at the end of this Contract if you accept its terms, the downloading of the manual VAISALA GMW115 will begin.