User manual LEICA MZ 10 F
DON'T FORGET : ALWAYS READ THE USER GUIDE BEFORE BUYING !!!
If this document matches the user guide, instructions manual or user manual, feature sets, schematics you are looking for, download it now. Diplodocs provides you a fast and easy access to the user manual LEICA MZ 10 F. We hope that this LEICA MZ 10 F user guide will be useful to you.
You may also download the following manuals related to this product:
LEICA MZ 10 F (1461 ko)
LEICA MZ 10 F (1475 ko)
LEICA MZ 10 F (1449 ko)
LEICA MZ 10 F (1452 ko)
LEICA MZ 10 F (1450 ko)
LEICA MZ 10 F (1451 ko)
LEICA MZ 10 F (1450 ko)
LEICA MZ 10 F BROCHURE (311 ko)
LEICA MZ 10 F BROCHURE (309 ko)
LEICA MZ 10 F BROCHURE (308 ko)
LEICA MZ 10 F BROCHURE (309 ko)
LEICA MZ 10 F BROCHURE (308 ko)
Manual abstract: user guide LEICA MZ 10 F
Detailed instructions for use are in the User's Guide.
[. . . ] Do not fit different plugs or cables Do not dismantle or modify components unless instructions for doing so are given in the user manual Components may only be opened by authorized personnel .
Stereo-fluorescence module
The stereo-fluorescence modul can be combined with the Leica stereomicroscope models MS5, MZ6, MZ7 . 5, MZ9 . 5, MZ12 . 5, MZ16, and with older models .
User manual
The present user manual describes only the functions of the fluorescence illuminator, the use of the filter sets, and the fitting of the fluorescence module . For detailed information about the use of the stereomicroscope, its care and its safety, please refer to those sections of the separate manual M2-105-0EN which relate to the Leica MZ12 . 5 with 12 . 5:1 zoom .
Place of use
Leica stereo-fluorescence systems are intended for use in closed rooms and may not be used outdoors .
Responsibilities of person in charge of instrument
Ensure that the Leica stereo-fluorescence systems are operated, maintained and repaired only by authorized and trained personnel Ensure that personnel who use the instrument have read and understood this user manual and in particular all safety instructions .
4
Servicing and repairs
Repairs may only be carried out by Leica-trained service technicians or by technical personnel authorized to do so by the person in charge of the instrument Only original Leica spare parts may be used Unplug the power cable before opening the supply unit . Touching live circuits can cause injury .
Supply unit
· Pull the power plug of the supply unit out from the power sokket:
when assembling and dismantling the lamp housing before opening the lamp housing when changing the mercury-vapour burner and other
components such as the heat-protection filter or the collector
when servicing the supply unit
Legal requirements
Adhere to general and local regulations relating to accident prevention and environmental protection .
Lamp housing
· Never open the lamp housing while the burner is switched on
(risk of explosion, UV irradiation, dazzling)
Conformity with European Community directive
Leica stereo-fluorescence systems and their accessories are constructed in accordance with the latest technologies and are provided with a statement of conformity with EC requirements .
· Always allow the lamp housing to cool for at least 15 minutes
before opening it (risk of burner exploding)
· Never cover the ventilator slots on the lamp housing (risk of fire)
Mercury-vapour lamp
· Read the user manual and safety directions provided by the
manufacturer of the mercury-vapour burner, and particularly those relating to its breakage and to the release of mercury
Light source: Safety regulations
Safety measures introduced by manufacturer
UV protection screen in front of the specimen plane prevents direct UV radiation from reaching the eyes . Dummy filter carriers in the unoccupied positions of the rapid filter changer prevent direct UV radiation from reaching the eyes UV filters in the observation beam paths protect the eyes against UV radiation . [. . . ] Connect the power cable to the supply unit and to the grid Make a note of the running-time meter reading .
1
4
25
3
Type Input voltage Inputfrequency Protection class Series no . Fuses
ebq 100dc 90V to 250V AC, max 265VA 48Hzto63Hz IP20 G 34240090
The fuse holder is secured in the holder with spring clips on both sides . Insert a screwdriver behind the spring clips and extract the fuse holder . 14 Refit the fuse holder
Adjusting the mercury vapour lamp This adjustment is important for obtaining a uniform light spot and good-quality fluorescence . While adjusting the mirror image, do not project light on to the electrodes for long periods (risk of explosion through overheating) The two electrodes are imaged as an extension of the plane of symmetry of the discharge arc, but are difficult to see . Switch on the supply unit and wait for two or three minutes · The100Wsupplyunitemitsawhistle Swing out the UV-light excluder (page 21) · WorkwithouttheUV-lightexcluderisonlypermittedforadjustment, and then only for short periods Pull out the light stop (see page 16, MZ10 F and page 22, Fluorescence module) . Mark a cross on a piece of paper and place it in the middle of the illuminated spot Select the lowest magnification
· Thedischargearcisvisibleintheilluminat dspotandlies e at about 45° to the lines of the cross (fig . a) Look into the eyepieces and focus on the cross Using your unaided eye, observe the discharge arc on the paper and bring it into focus with the focusing knob (3) (fig . a) Using the positioning knobs (1 and 2), displace the discharge arc (fig . b) Bring the mirror image of the discharge arc into focus with the knob (6) and use the knobs 4 and 5 to position it symmetrically relative to the original image (fig . c) · Forthe50Wburner, thedischarge-arcmageandits i mirror image should touch · Forthe100Wburner, thedischarge-arcmageanditsmirror i image should be superimposed . Using the focusing knob (3), readjust the illuminated field · Theilluminatedfieldshouldnowbelarge, circularand as uniform as possible Reposition the UV-light excluder correctly (page 21) .
Hg50W
Hg100W
3
Figs . c: Focusing and positioning the mirror image of the discharge arc
4 5 6
15
Leica MZ10 F
Dummy filter carrier In the dummy filter carrier there are two openings for the observation beam paths instead of the barrier filters . Use the dummy filter carrier: if you are using less than four sets of filters if you are working without the fluorescence illuminator for periods of up to 15 seconds . · Donotusethedummyfiltercarriertoblockthelightfromthe mercury-vapour burner for more than 15 seconds, or it will heat up . Light stop The FLUOIII filter system includes a slot for the light stop . transmitted light instead of under conditions of fluorescence, the light stop is used to block the illumination beam path so that the mercuryvapour burner does not have its life shortened by being switched on and off more than is absolutely necessary (see page 5) .
FLUOIII filter system
The filter system consists of a rapid filter changer for the barrier- and excitation filters and two slots for a light stop and for a filter slide respectively . Rapid filter changer The rapid filter changer of the Leica MZ10 F accepts a total of four sets of filters . Each filter set consists of two barrier filters for the observation beam paths and one excitation filter for the illumination beam path . To avoid fingerprints on the filters in the filter sets, try to avoid touching them . If fingerprints do occur, remove them with a clean lint-free cloth and pure alcohol . An empty filter carrier is available for individual filter combinations .
Insert a set of filters into the rapid filter changer so that the inscription (e . g . [. . . ] For the 100W lamp, wait two or three minutes for the whistle signal .
23
Leica Microsystems the brand for outstanding products
Leica Microsystems' mission is to be the world's first-choice provider of innovative solutions to our customers' needs for vision, measurement and analysis of micro-structures . Leica, the leading brand for microscopes and scientific instruments, developed from five brand names, all with a long tradition: Wild, Leitz, Reichert, Jung and Cambridge Instruments . The companies of the Leica Microsystems Group operate internationally in three business segments, where we rank with the market leaders .
· Microscopy Systems
Leica Microsystems an international company with a strong network of customer services.
Australia: Gladesville Austria: Vienna Canada: RichmondHill/Ontario Denmark: Herlev France: Rueil-Malmaison Germany: Bensheim Italy: Milan Japan: Tokyo Korea: Seoul Netherlands: Rijswijk People'sRep. ofChina:HongKong Portugal: Lisbon Singapore Spain: Barcelona Sweden: Sollentuna Switzerland: Heerbrugg United Kingdom: Milton Keynes USA: Bannockburn/lllinois Tel . +1 847 405 0123 Fax +61 2 9817 8358 Fax +43 1 486 80 50 30 Fax+19057628937 Fax+4544540111 Fax +33 1 47 32 85 86 Fax +49 6251 136 155 Fax +39 0257 40 3475 Fax +81 3 5421 2896 Fax +82 2 514 65 48 Fax +31 70 4132 109 Fax+85225644163 Fax +351 21 385 4668 Fax +65 6773 0628 Fax +34 93 494 95 32 Fax +46 8 625 45 10 Fax+41717263444 Fax +44 1908 609 992 Fax +1 847 405 0164